About the NETDAQ-2640A

The Fluke 2640A NetDAQ is a 20-channel networked data acquisition unit. Together with Fluke's NetDAQ Logger software on a Windows PC it forms a data logging system that scans the analog channels in sequence and calculates up to ten additional computed channels from the readings.

It measures DC and AC volts, resistance, frequency, DC current (with a shunt) and temperature with thermocouples (types B, C, E, J, K, N, R, S, T) or RTDs. The 2640A is the precision model of the pair: 5 1/2 digits, around 0.02% DC accuracy and 150 V common-mode voltage (300 V on channels 1 and 11), with up to 143 readings per second; the otherwise identical 2645A trades accuracy for up to 1,000 readings per second.

Wiring goes to a removable Universal Input Module (2620A-100 connector set), and the rear panel carries digital I/O, totalizer, alarm/trigger I/O, RS-232 and Ethernet. Scans can be triggered by interval, alarm or an external signal; one PC can control up to 20 instruments. It runs on AC mains or 9 to 16 V DC and can be rack-mounted.

Fluke lists the 2640A and 2645A NetDAQ as discontinued and its page names no direct successor.

Frequently asked questions

What is the difference between the 2640A and 2645A?

They are identical in operation; the 2640A emphasises precision (5 1/2 digits, about 0.02% accuracy, up to 143 readings/s), the 2645A speed (4 1/2 digits, about 0.04%, up to 1,000 readings/s).

Which software runs it?

NetDAQ Logger for Windows (2640A-911) or NetDAQ Logger with Trending (2640A-912); Fluke DAQ 6.0.4 also supports the 2640A, but Fluke lists that software as discontinued too.

How does it connect to the PC?

Over Ethernet: 10BaseT twisted-pair or 10Base2 coax; Fluke's 2005 supplement notes that later units ship without the 10Base2 port.

Is it still made?

No. Fluke lists the 2640A and 2645A NetDAQ as discontinued.

Documents

Specifications

Power 107 to 264 V ac 50/60 Hz, or 9 to 16 V dc
Network Ethernet 10BaseT and 10Base2 (later units: 10BaseT only)
Resolution 5 1/2 digits
Scan speed Up to 143 readings/s (fast, 20 channels); 6/s slow
Serial port RS-232C, 9-pin
Analog channels 20
Size and weight 9.3 x 21.6 x 36.2 cm (H x W x D), 4 kg
Basic DC accuracy About 0.02% (90 mV to 3 V ranges, 0.01% of input + offset, 90 days, slow rate)
Computed channels 10
Thermocouple types B, C, E, J, K, N, R, S, T
Common mode voltage 150 V (300 V on channels 1 and 11)
Measurement functions DC volts, AC volts, ohms, frequency, DC current (with shunt), thermocouple and RTD temperature
Operating temperature -20 to 60 °C
Digital I/O and alarm lines 12 I/O lines total, plus totalizer and master alarm output